Por favor, use este identificador para citar o enlazar este ítem: http://dspace.espoch.edu.ec/handle/123456789/8603
Registro completo de metadatos
Campo DC Valor Lengua/Idioma
dc.contributor.advisorBorja Saavedra, Myrian-
dc.contributor.authorPaguay Villarroel, Jessica Cristina-
dc.date.accessioned2018-08-06T20:42:32Z-
dc.date.available2018-08-06T20:42:32Z-
dc.date.issued2018-06-
dc.identifier.citationPaguay Villarroel, Jessica Cristina. (2018). Verificación del comportamiento dosimétrico del sistema de planificación de tratamientos de radioterapia a través de las curvas de calibración de densidades electrónicas relativas (DER) en el Hospital de la sociedad de lucha contra el cáncer (SOLCA) núcleo Loja. Escuela Superior Politécnica de Chimborazo. Riobamba.es_ES
dc.identifier.urihttp://dspace.espoch.edu.ec/handle/123456789/8603-
dc.descriptionEl presente trabajo persigue como objetivo la verificación del comportamiento dosimétrico del Sistema de Planificación de Tratamientos de Radioterapia a través de las curvas de calibración de Densidades Electrónicas Relativas (DER) en el Hospital de la Sociedad de Lucha Contra el Cáncer (SOLCA) Núcleo Loja realizada en el periodo octubre 2017-febrero 2018. Para ello se utilizó un fantoma antropomorfo CIRS 062M y un tomógrafo Toshiba Activion 16 para la obtención de las Unidades Hounsfield y se tomaron los valores de densidades electrónicas especificadas en el manual de dicho fantoma para la obtención de la nueva curva de calibración DER. Para la realización de los controles de calidad dosimétricos y la planificación clínica de tres casos en localización de pulmón, pelvis y cráneo se utilizó un sistema de planificación Eclipse 13, CLINAC CX, y una cámara de ionización PTW tipo Farmer con un volumen sensible de 0.6 cm3 conectada a un electrómetro PTW UNIDOS E. Los resultados muestran que las mediciones realizadas en cada uno de los insertos del fantoma antropomorfo en ningún caso excedieron los límites establecidos de tanto en el tomógrafo como en el TPS Eclipse 13, las curvas de calibración DER tuvieron comportamientos similares en todos los tipos de insertos estudiados excepto en hueso, las pruebas de control de calidad recomendadas no superaron el límite máximo de tolerancia establecida por la OIEA del ±4%, y las tres planificaciones clínicas en tórax, pelvis y cráneo, sus desviaciones en el cálculo de las dosis absorbida por el TPS y la obtenida por medición no supera la tolerancia del ±5% planteada porla ICRU.es_ES
dc.description.abstractThis work aims to verify the dosimetric accuracies of Eclipse (Version 13) Treatment Planning System (TPS) using calibration curves of relative electron densities (RED) at SOLCA (Hospital de la Sociedad de Lucha Contra el Cáncer) in Loja during the period of October 2017 – February 2018. A CIRS 062M anthropomorphic phantom and a Toshiba Activion 16 CT Scan were used to obtain Hounsfield Units, and the specific electron density values were taken in the manual of that phantom to obtain the new RED calibration curve. For the dosimetric quality controls and the clinical planning of chest, pelvis, and skull cases, an Eclipse version 13 planning system, CLINAC CX, and a PTW model Farmer-type ionization chamber (sensitive volume of 0.6 cm3) connected to a PTW UNIDOS E electrometer were used. The results showed that measures taken in each one of the anthropomorphic phantom inserts did not exceed ±20 HU using the CT scan and TPS (Eclipse Version 13). The RED calibration curves showedsimilar behaviors in all types of studied inserts, but in bone one. The recommended quality control tests did not exceed the maximum limit of tolerance of ±4% established by International Atomic Energy Agency (IAEA), and the chest, pelvis and skull clinical planning, their deviations in the calculation of the absorbed-dose by TPS and the calculated dose did not exceed the tolerance of ±5% established by ICRU (International Commission on Radiation Units and Measurements).es_ES
dc.language.isospaes_ES
dc.publisherEscuela Superior Politécnica de Chimborazoes_ES
dc.relation.ispartofseriesUDCTFC;86T00062-
dc.subjectBIOFÍSICAes_ES
dc.subjectDENSIDAD ELECTRÓNICA RELATIVA (DER)es_ES
dc.subjectSISTEMA DE PLANIFICACIÓN (TPS)es_ES
dc.subjectUNIDADES HOUNSFIELDes_ES
dc.subjectDOSIS PLANIFICADAes_ES
dc.subjectDOSIS ENTREGADAes_ES
dc.subjectFANTOMA ANTROPOMORFOes_ES
dc.titleVerificación del comportamiento dosimétrico del sistema de planificación de tratamientos de radioterapia a través de las curvas de calibración de densidades electrónicas relativas (DER) en el Hospital de la sociedad de lucha contra el cáncer (SOLCA) núcleo Loja.es_ES
dc.typebachelorThesises_ES
dc.contributor.miembrotribunalGonzales, Nadia Aimé-
Aparece en las colecciones: Biofísica

Ficheros en este ítem:
Fichero Descripción Tamaño Formato  
86T00062.pdf3,31 MBAdobe PDFVista previa
Visualizar/Abrir


Los ítems de DSpace están protegidos por copyright, con todos los derechos reservados, a menos que se indique lo contrario.