Por favor, use este identificador para citar o enlazar este ítem: http://dspace.espoch.edu.ec/handle/123456789/13251
Registro completo de metadatos
Campo DC Valor Lengua/Idioma
dc.contributor.advisorYanchapanta Bastidas, Vilma Nohemí-
dc.contributor.authorVásconez Zabala, Cristian Aníbal-
dc.date.accessioned2019-12-06T15:00:12Z-
dc.date.available2019-12-06T15:00:12Z-
dc.date.issued2019-10-29-
dc.identifier.citationVásconez Zabala, Cristian Aníbal. (2019). Validación del blindaje estructural de las salas de radiodiagnóstico e intervencionismo del Hospital de Especialidades Carlos Andrade Marín mediante levantamiento radiométrico bajo el protocolo IAEA/ARCAL XLIX. Escuela Superior Politécnica de Chimborazo. Riobamba.es_ES
dc.identifier.urihttp://dspace.espoch.edu.ec/handle/123456789/13251-
dc.descriptionEl objetivo principal es validar el blindaje estructural en las salas de radiodiagnóstico e intervencionismo del Hospital de Especialidades Carlos Andrade Marín, mediante un levantamiento radiométrico bajo el protocolo IAEA/ARCAL XLIX. Para lo cual, se replicó los exámenes más comunes realizados en las diferentes salas con la ayuda de un fantoma antropomórfico que simula al paciente. Se tomaron distintos puntos de interés en cada sala donde se encuentre ubicado el Personal Ocupacionalmente Expuesto (POE) y el público. Las diferentes salas cuentan con equipos de rayos x convencional, fluoroscopia, tomografía computarizada y una sala de angiografía con equipo arco en C. Se tomaron medidas de tasa de dosis con un detector gaseoso tipo cámara de ionización modelo Radcal RSM, a diferentes distancias según sea paredes, piso o techos, para posteriormente estimar la dosis semanal y anual considerando variables como voltaje, corriente, factor de uso, factor de calibración de la cámara de ionización, factor de ocupación, carga de trabajo semanal y la lectura obtenida. Se concluye que las dosis semanales y anuales en los distintos puntos de interés de cada sala, cumple con los niveles de restricción de dosis permitidos de 0.10 mSv/sem o 5.0 mSv/año para zonas controladas y 0,01 mSv/sem o 0.5 mSv/año para zonas no controladas. Se recomienda mantener la distribución de las salas para no alterar la clasificación de las zonas y contar siempre con los equipos calibrados.es_ES
dc.description.abstractThe main objective is to validate the structural shielding in the radiodiagnostic and intervention rooms of the Carlos Andrade Marín Specialty Hospital, by means of a radiometric survey under the IAEA / ARCAL XLIX protocol. . For which, the most common examinations performed in the different rooms were repeated with the help of an anthropomorphic phantom that simulates the patient. Different points of interest were taken in each room where the Occupationally Exposed Personnel (OEP) and the public are located. The different rooms have conventional x-ray equipment, fluoroscopy, computed tomography and an angiography room with C-arc equipment. Dose rate measurements were taken with a gaseous detector type Radcal RSM ionization chamber, at different distances as walls, floor or ceilings, to later estimate the weekly and annual dose considering variables such as voltage, current, use factor, ionization chamber calibration factor, occupancy factor, weekly workload and the reading obtained. It is concluded that the weekly and annual doses at the different points of interest of each room comply with the permitted dose restriction levels of 0.10 mSv / wk or 5.0 mSv / year for controlled areas and 0.01 mSv / wk or 0.5 mSv / year for uncontrolled areas. It is recommended to maintain the distribution of the rooms so as not to alter the classification of the zones and always have the calibrated equipment.es_ES
dc.language.isospaes_ES
dc.publisherEscuela Superior Politécnica de Chimborazoes_ES
dc.relation.ispartofseriesUDCTFC;86T00083-
dc.rightsinfo:eu-repo/semantics/openAccesses_ES
dc.subjectBIOFÍSICAes_ES
dc.subjectRADIOLOGÍAes_ES
dc.subjectLEVANTAMIENTO RADIOMÉTRICOes_ES
dc.subjectPROTECCIÓN RADIOLÓGICAes_ES
dc.subjectTASA DE DOSISes_ES
dc.subjectZONA CONTROLADAes_ES
dc.subjectZONA SUPERVISADAes_ES
dc.subjectPROTOCOLO ARCALes_ES
dc.subjectFANTOMA ANTROPOMÓRFICOes_ES
dc.titleValidación del blindaje estructural de las salas de radiodiagnóstico e intervencionismo del Hospital de Especialidades Carlos Andrade Marín mediante levantamiento radiométrico bajo el protocolo IAEA/ARCAL XLIX.es_ES
dc.typebachelorThesises_ES
dc.contributor.miembrotribunalHeredia Moyano, María Fernanda-
dc.rights.urihttps://creativecommons.org/licenses/by-nc-sa/3.0/ec/es_ES