dc.identifier.citationArias Garcés, Mario Javier. (2014). Comisionmiento y puesta en marcha del acelerador lineal digital de alta energía Elekta Synergy para rayos X en el departamento de radioterapia del hospital Carlos Andrade Marín. Escuela Superior Politécnica de Chimborazo. Riobamba.es_ES
dc.identifier.urihttp://dspace.espoch.edu.ec/handle/123456789/3431-
dc.descriptionEl comisionamiento y puesta en marcha para Rayos X del acelerador lineal Elekta Synergy se realizó en el Departamento de Radioterapia del Hospital Carlos Andrade Marín (HCAM) de la ciudad de Quito. Con la ayuda del Método Deductivo partimos de conceptos generales en Física Médica hasta llegar a conceptos específicos como dosis absorbida, magnitudes absolutas y relativas de radioterapia, el Método Inductivo; de razonamiento, contribuyó a formular las conclusiones, mediante la Técnica de la Observación Estructurada y Bibliográfica, se recopiló la información necesaria, se comprendió y aplicó la guía Beam Data Collect List, reporte Task Group - 106 de la Asociación Americana de Físicos Médicos (AAPM), Technical Report Series (TRS) - 398 y TRS - 430 de la Organización Internacional de Energía Atómica (IAEA). Se utilizaron detectores de radiación, fantomas y electrómetros para realizar las mediciones, mediante el protocolo TG - 106 se tuvo en cuenta la forma correcta de posicionar los detectores dentro del fantoma y varios factores como sensibilidad y respuesta de detectores, radiación de fuga, velocidad de escaneo, entre otros. Posteriormente se calibró el acelerador utilizando el reporte TRS - 398, obteniendo 1.0016 cGy/MU con 6 MV y 1.0034 cGy/MU con 15 MV y se verificaron las unidades monitor (MU) calculadas por el Sistema de Planificación y dosis medida in vivo, los resultados obtenidos están dentro de los límites de tolerancia recomendados por el reporte TRS - 430. Se concluye que el acelerador lineal se encuentra listo para su uso clínico debido a que cumple con todos los parámetros recomendados por los protocolos internacionales, se recomienda utilizar los resultados de este trabajo como referencia si se requiere realizar un re - comisionamiento del equipo.es_ES
dc.description.abstractThe commissioning and start-up of the Elekta Synergy linear accelerator for X-rays was carried out at the Radiotherapy Department of the Carlos Andrade Marín Hospital (HCAM) in the city of Quito. With the help of the Deductive Method we started from general concepts in Medical Physics to specific concepts such as absorbed dose, absolute and relative magnitudes of radiotherapy, the Inductive Method; The Inductive Method of reasoning contributed to formulate the conclusions, through the Structured and Bibliographic Observation Technique, the necessary information was collected, the Beam Data Collect List guide, Task Group - 106 report of the American Association of Medical Physicists (AAPM), Technical Report Series (TRS) - 398 and TRS - 430 of the International Atomic Energy Agency (IAEA) were understood and applied. Radiation detectors, phantoms and electrometers were used to perform the measurements, using the TG - 106 protocol, taking into account the correct way to position the detectors inside the phantom and several factors such as sensitivity and response of detectors, leakage radiation, scanning speed, among others. Subsequently, the accelerator was calibrated using the TRS - 398 report, obtaining 1.0016 cGy/MU with 6 MV and 1.0034 cGy/MU with 15 MV and the monitor units (MU) calculated by the Planning System and dose measured in vivo were verified; the results obtained are within the tolerance limits recommended by the TRS - 430 report. It is concluded that the linear accelerator is ready for clinical use because it complies with all the parameters recommended by international protocols, it is recommended to use the results of this work as a reference if a re-commissioning of the equipment is required.es_ES
dc.language.isospaes_ES
dc.publisherEscuela Superior Politécnica de Chimborazoes_ES
dc.relation.ispartofseriesUDCTFC;86T00016-
dc.rightsinfo:eu-repo/semantics/openAccesses_ES
dc.subjectBIOFÍSICAes_ES
dc.subjectMEDICINAes_ES
dc.subjectDETECTORES DE RADIACIÓNes_ES
dc.subjectCONTROL DE CALIDADes_ES
dc.subjectRADIOTERAPIAes_ES
dc.subjectTRATAMIENTO CLÍNICOes_ES
dc.subjectRAYO Xes_ES
dc.subjectDOSIMETRIA [RADIACIÓN]es_ES
dc.subjectACELERADOR LINEAL CLÍNICO [Instalación]es_ES
dc.subjectACELERADOR LINEAL DIGITAL [Instalación]es_ES
dc.subjectHOSPITAL CARLOS ANDRADE MARINes_ES
dc.subjectQUITO (DISTRITO METROPOLITANO)es_ES
dc.subjectPROVINCIA DE PICHINCHAes_ES
dc.titleComisionmiento y puesta en marcha del acelerador lineal digital de alta energía Elekta Synergy para rayos X en el departamento de radioterapia del hospital Carlos Andrade Marínes_ES
dc.typebachelorThesises_ES
dc.contributor.miembrotribunalOrbe Ordoñez, Jheny del Carmen-
dc.rights.urihttps://creativecommons.org/licenses/by-nc-sa/3.0/ec/es_ES
Aparece en las colecciones: Biofísica

Ficheros en este ítem:
Fichero Descripción Tamaño Formato  
86T00016.pdf7,93 MBAdobe PDFVista previa
Visualizar/Abrir


Este ítem está sujeto a una licencia Creative Commons Licencia Creative Commons Creative Commons